BÜCHI Labortechnik AG Editing a method | 8
Operation Manual FatExtractor E-500 57/88
Open time Level sensor posi-
tion
Explanation
[mid] middle The sample structure is easily releasing
the solvent.
[long] top The sample structure is easily releasing
the solvent.
bottom The sample structure is slowly releasing
the solvent.
[max] top The Soxhlet LSV chamber is used.
The sample structure is slowly releasing
the solvent.
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [Sox valve open time] with the navigation control.
ð The display shows a dialog with the selectable values.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The Soxhlet valve open time is saved.
8.8 Setting the rinse time of a method (Soxhlet and Hot Extraction
only)
Enter the duration of the extraction step in minutes.
at least 0 minutes
maximum 5940 minutes / 99 hours
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [Rinse time] with the navigation control.
u Turn the navigation control until the required value.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The time is saved.
8.9 Setting the rinse heating level of a method (Soxhlet and Hot
Extraction only)
The default setting for the rinse heating level is depending on the selected solvent.
Change the rinse heating level to compensate ambient conditions.