BÜCHI Labortechnik AG Operating the interface pro | 7
Operation Manual FatExtractor E-500 51/88
7.3.3 Configuration menu
On the [Configuration] menu, you can enter a variety of settings and retrieve informa-
tion.
Submenu Settings
The submenu [Settings] contains system settings for the instrument.
Action Option Explanation
[Language] Choice of display lan-
guage on the interface
The following languages are avail-
able:
English/German/French/Italian/
Spanish/Japanese/Chinese/Russian/
Polish
[Date] Date input Enter in sequence: Day, month,
year. Apply the settings by pressing
[OK].
[Time] Time input Enter in sequence: Minutes, hours.
Apply the settings by pressing [OK].
[Beeper] Off/On Setting for audible signal in re-
sponse to input controls.
[Display brightness] Enter setting Display brightness level in %: 0 -
100
[Configuration] Choice of extraction
method
The following configurations are
available:
SOX (Soxhlet Extraction) / HE (Hot
Extraction) / ECE (Economic Contin-
uous Extraction)
[Max. Cycle time]
(Soxhlet Extraction
only)
Enter value Maximum cycle time in minutes: 0 -
240
[Analyte protection
(option)]
View Threshold for empty-no beaker
Threshold for full-empty beaker
[Mobile connection
QR code]
View The control unit shows a QR code
for connecting the Extraction Re-
ports app with the instrument.
[Mobile connection
password]
View The control panel shows a password
for entry in the Extraction Reports
app.
[Network] Enter value The following parameters can be
edited:
Device name/DHCP/MAC address/
System IP address/Subnet mask/
Gateway
[Delete app con-
nection]
Confirmation question Resets external connections to the
instrument.
[Demo Mode] On/Off Simulate an extraction.