BÜCHI Labortechnik AG Editing a method | 8
Operation Manual FatExtractor E-500 59/88
The following parameters are set automatically:
Rinse heating level (without analyte protection only)
Drying time
Drying heating level
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Tap the action [SmartDrying].
ð The function is enabled.
8.12 Setting the drying time of a method
NOTICE
Risk of beaker damage or deterioration of the analyte due to incorrect
settings of the drying step.
Incorrect drying step settings can damage the analyte and the beaker.
u Changing the default settings is at your own risk.
The duration of the drying time of a method depends on the solvent volume.
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [Drying time] with the navigation control.
u Turn the navigation control until the required value.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The time is saved.
8.13 Setting the drying heating level of a method
The default setting for the drying heating level is depending on the selected solvent.
Change the drying heating level to compensate fluctuations in ambient conditions.