8 | Editing a method BÜCHI Labortechnik AG
60/88 Operation Manual FatExtractor E-500
WARNING
Setting the heating level too high
The consequences are device damage, serious injuries or death.
u Use the pre-set heating levels.
u Adjust the heating level to compensate the ambient conditions.
Navigation path
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [Drying heating level] with the navigation control.
u Turn the navigation control until the required value.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The heating level is saved.