8 | Editing a method BÜCHI Labortechnik AG
58/88 Operation Manual FatExtractor E-500
WARNING
Setting the heating level too high
The consequences are device damage, serious injuries or death.
u Use the pre-set heating levels.
u Adjust the heating level to compensate the ambient conditions.
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [Rinse heating level] with the navigation control.
u Turn the navigation control until the required value.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The heating level is saved.
8.10 Setting the number of drains for a method (Hot Extraction only)
During each drain 3 - 4 mL of solvent is drained into the solvent tank.
The level in the beaker is lowered to finally rinse the sample without touching the ex-
tract.
Navigation path
➔
u Navigate to the [Method] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [Edit] on the function bar.
u Select the action [No. of drains] with the navigation control.
ð The display shows a dialog with the selectable values.
u Select the value you wish to use with the navigation control.
u Tap the function [OK] on the function bar.
u Tap the function [Save] on the function bar.
ð The value is saved.
ð The dialog box closes.
8.11 Setting SmartDrying (Soxhlet and Economic Continuous
Extraction only)
Smart-Drying automatically sets the parameters for the drying step.