BÜCHI Labortechnik AG Cleaning and servicing | 11
Operation Manual FatExtractor E-500 75/88
u Push the solvent tank up.
u Tilt the solvent tank at 20° and remove it.
11.3 Draining solvent from the instrument
1. Switch on the pump. See Chapter11.3.1 "Switch on the pump", page75
2. Open the tank valves. See Chapter11.3.2 "Open the tank valves", page75
3. Switch of the valves. See Chapter11.3.3 "Switch of the pump", page76
11.3.1 Switch on the pump
Navigation path
[Service] [Actors]
Precondition:
R A tank bottle is installed. See Chapter5.8 "Preparing the tank bottle", page30
u Navigate to [Actors] via the navigation path.
ð The display shows the view Actors.
u Enable the action [Pump].
11.3.2 Open the tank valves
Navigation path
[Service] [Actors]
[Valves]
u Navigate to [Valves] via the navigation path.
ð The display shows the view Valves.
u Select the position of the tank valve you wish to open.
u Open the tank valve.
u Wait until all solvent is in the tank bottle.