BÜCHI Labortechnik AG Cleaning and servicing | 11
Operation Manual FatExtractor E-500 75/88
u Push the solvent tank up.
u Tilt the solvent tank at 20° and remove it.
11.3 Draining solvent from the instrument
1. Switch on the pump. See Chapter11.3.1 "Switch on the pump", page75
2. Open the tank valves. See Chapter11.3.2 "Open the tank valves", page75
3. Switch of the valves. See Chapter11.3.3 "Switch of the pump", page76
11.3.1 Switch on the pump
Navigation path
➔ ➔ [Service] ➔ [Actors]
Precondition:
R A tank bottle is installed. See Chapter5.8 "Preparing the tank bottle", page30
u Navigate to [Actors] via the navigation path.
ð The display shows the view Actors.
u Enable the action [Pump].
11.3.2 Open the tank valves
Navigation path
➔ ➔ [Service] ➔ [Actors] ➔
[Valves]
u Navigate to [Valves] via the navigation path.
ð The display shows the view Valves.
u Select the position of the tank valve you wish to open.
u Open the tank valve.
u Wait until all solvent is in the tank bottle.