EasyManua.ls Logo

burmeier Dali - Page 27

burmeier Dali
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sheathing
Phenomenon EMC basic standard
or test method
Immunity level (test + compli-
ance)
Home healthcare environment
Electrostatic discharge (ESD) IEC 61000-4-2 +/- 8 kV contact
+/- 2 kV, +/- 4 kV¸ +/- 8 kV, +/- 15 kV;
+/- 25 kV air
High-frequency electromagnetic
fields
IEC 61000-4-3 10 V/m ; (80 MHz to 2.7 GHz; 80%
AM at 1 kHz)
High-frequency electromagnetic
fields in the immediate vicinity of
wireless communication devices
IEC 61000-4-3 See separate table zz (at the end of
this chapter)
Magnetic fields with rated power
frequencies
IEC 61000-4-8 See separate table zz (at the end of
this chapter)
AC port for supply input
Phenomenon EMC basic standard Immunity level (test + compli-
ance)
Home healthcare environment
Electrical fast transient distur-
bances/bursts
IEC 61000-4-4 +/- 2 kV; 100 kHz repetition frequen-
cy
Electrical surges: conductor to
conductor
IEC 61000-4-5 +/- 0,5 kV; +/- 1kV
Conducted interference induced
by high-frequency fields
IEC 61000-4-6 3 V; 0.15 MHz to 80 MHz; 6V in ISM
and amateur radio frequency bands
between 0.15 MHz and 80MHz 80%
AM at 1 kHz
Voltage dips IEC 61000-4-11 0% UT; 1/2 period; at 0, 45, 90, 135,
180, 225, 270 and 315 degrees
0% UT; 1 period; and 70% UT; 25
periods; single-phase at 0 degrees
Celsius
Voltage interruptions IEC 61000-4-11 0% UT; 250 periods
Product Description
Part A: General Information 21

Table of Contents