Ports for signal input/signal output parts
Phenomenon EMC basic standard Immunity level (test + compli-
ance)
Home healthcare environment
Electrostatic discharge (ESD) IEC 61000-4-2 +/- 8 kV; contact
+/- 2 kV, +/- 4 kV¸ +/- 8 kV, +/- 15 kV;
+/- 25kV air;
Electrical fast transient distur-
bances/bursts
IEC 61000-4-4 +/- 1 kV; 100 kHz repetition frequency
Conducted interference in-
duced by high-frequency
fields
IEC 61000-4-6 3 V; 0.15 MHz to 80 MHz; 6V in ISM
and amateur radio frequency bands
between 0.15 MHz and 80MHz 80%
AM at 1 kHz
Table zz: Test specifications for the immunity of sheathings to high-frequency wire-
less communication equipment
Test fre-
quency
MHz
Frequen-
cy band
Radio serv-
ice
Modulation Max.
power W
Distance
m
Immunity
test level
v/m
385 380 to 390 TETRA 400 Pulse modu-
lation 18 Hz
1.8 0.3 27
450 430 to 470 GMRS 460
FRS460
FM +/- 5%
deviation,
1kHz sine
wave
2 0.3 28
710 704 to 787 LTE band 13,
17
Pulse modu-
lation 217 Hz
0.2 0.3 28
745
780
810 800 to 960 GSM
800/900 TET-
RA 800
iDEN820,
CDMA 850,
LTE band 5
Pulse modu-
lation 18 Hz
0.2 0.3 28
870
930
1720 1700 to
1990
GSM 1800
CDMA 1900,
Pulse modu-
lation 18 Hz
2 0.3 28