Inspection report
Result: Bed prot.
class II (type BF)
≥ 70 MΩ MΩ
Leakage current (direct or differential current measure-
ment) (type BF)
Proceed as follows:
OK Not
OK
Description of de-
fect
1. Plug the mains cable/switch mode power supply into
the test socket of the measuring instrument.
2. Connect the probe of the measuring instrument to
the bed; measuring point: Bare metal screw under
backrest in frame of mattress base
3. Operate the motors using the handset for the dura-
tion of the measurement
4. Start the measurement procedure on the measuring
instrument.
Limit value Current value
(normalised to
rated value of
mains voltage)
Result: Bed prot.
class II (type BF)
0.1 mA mA
In case of measured voltage external
conductor - earth
volt:
Inspection report
III. Functional check OK Not
OK
Description of
defect
Functional check of the electrical components
What? How?
End of travel cut-out of the motors Automatic cut-out
Pairing the Bluetooth handset with
the control unit
Perform the test acc. to
Part B: Pairing the Blue-
tooth Handset » 25
External power supply/handset No ‘rattling’ when shaken?
Maintenance
Part B: Operator and Technical Personnel 37