Inspection report
Handset: Operating function, lock-
ing function
Perform the test acc. to
Part C: Handset » 9
Motors Abnormal noise develop-
ment (rattling, uneven run-
ning)
Strain relief of mains cable (if mains
cable available)
Mains cable firmly fas-
tened
Functional check of the mechanical components
Joints and pivots Smooth operation
Grab handle with strap Securely fixed when load
tested under approx. 75
kg load (hang from it brief-
ly with two hands)
Castors Moving and braking
Emergency release of the backrest Test according to instruc-
tion manual
Safety side Securely engaged, secure
position, unlocking
Lower leg rest Engages properly
Accessories (e.g. patient lifting
pole, grab handle)
Correct fastening, no dam-
age, suitable for purpose
Inspection result:
Inspection passed; test approval sticker applied:
[ ] Safety or functional defects were not detected
[ ] No direct risk, the defects detected can be rectified quickly
Inspection was not passed; no test approval sticker applied:
[ ] Device must be taken out of circulation until the defects have been rectified!
[ ] Device does not conform to requirements – modification/replacement of components/decommis-
sioning recommended!
All values within permissible range:[ ] yes [ ] no Next inspection date: