4 Operation
BodyGuard 595Color Vision Operator Manual
22
4 Operation
Before Operating the Pump
Before attaching the system to a patient, run the following tests to verify that all
indicators and alarms work properly.
When an alarm is activated the following occurs:
An alarm message appears on the display
An audible alarm sounds
Infusion ceases
The LED changes from green to red
NOTE: all tests should be performed in the Continuous program while the
pump is not connected to patient.
Pump Operation Test
1.
Click the pump into a Charger connected to AC power. Verify that the CHARGE
indicator is on, red or green light (on charger).
2.
Insert a MicroSet
TM
administration set and close the door. Press the door until a
click is heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
3.
Press until the Self-Test screen appears. If air sensor is off, press to
confirm.
4.
During the Self-Test a long beep will be heard. This procedure verifies that the
pump’s acoustic and visual features are working properly. After self test is
completed, the display will show the last rate setting.
5.
Prime the set according to the priming instructions following later in this manual.
6.
When priming is completed, wait 2 minutes
without pressing any key. After two minutes the
screen displays “Pump Paused Too Long and an
alarm will sound. Press to silence the alarm
again or to exit alarm condition.
REGIONAL 10:00 - 70%
Pump Paused
Too Long
Press key

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 595 Color Vision and is the answer not in the manual?

CME BodyGuard 595 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 595 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals