BodyGuard 595™Color Vision Operator Manual
The display screen will show a graph indicating the
Bolus is being infused. The graphical bar displays
the amount of the bolus to be infused (grey) and
the amount infused (green).
NOTE: once the bolus is completed, the pump will resume the original infusion
program.
NOTE: user can stop bolus at any time by pressing .
NOTE: clinician bolus can also be set using 'Clinician Bolus' option on the menu.
It is recommended to select a protocol prior to starting a clinician bolus.
Automatic Bolus Rate Adjustment
The BodyGuard 595 Color Vision™ has a special feature which optimizes bolus rate
during infusion through narrow catheters such as Epidural catheters. The pump will
attempt to deliver the bolus at the pre-set bolus delivery rate. If the pump senses back
pressure above the pre-set occlusion pressure level during bolus delivery it will
automatically reduce the bolus delivery speed until it senses a pressure below the level
at which an occlusion alarm would normally sound. If it still senses an occlusion
pressure above this level at the minimum bolus delivery rate the pump will then alarm
‘Down Occlusion’. This feature ensures optimum bolus delivery speed and reduces false
Occlusion alarms during bolus delivery through narrow catheters such as Epidural or
soaker catheters.
Interpreting the Display Screen During Infusion
EPIDURAL 10:00 - 70% █
■
Bolusing 0.363 mL
███████████████████
Bolus Dose 5 mL
2 mL/hr
██████████████████████
20 of 400 mmHg
REGIONAL 10:00 - 70% █
■
A: FENTANYLCITRATE
air
Rate :10.5 mL/hr
Patient Bolus Dose 1.5 mL
Lockout Time 20 Minutes
Auto-Bolus Dose 0.5 mL
Auto-Bolus interval 2 Minutes
███████████
█
█████████████
█
███████████
20 of 400 mmHg
Protocol &
Protocol Letter
Color coded
Basal Rate
(ml/hr or mg/ml)
Time between two patient
boluses (when applicable)
Only if air sensor
disable
Title bar, color coded
PCA (blue), Epidural
(Yellow), Regional (Green)
Battery Status Indicator,
Actual time, Type of
treatment
Patient bolus volume
(When applicable)
Time between two auto
boluses (when applicable)
Auto bolus volume
(When applicable)