BodyGuard 595™Color Vision Operator Manual
The BodyValve™
The Anti-Free Flow valve, BodyValve™, enhances pump functioning by:
Preventing free-flow in the event the set is detached from the pump.
Preventing back-flow (reflux) in the event several infusion pumps are connected
simultaneously to the same patient.
Preventing free-flow in the event of a mechanical malfunction.
Preventing pump operation if the set has been loaded incorrectly.
Warning: replace the administration set every 72 hours to lessen the
incidence of bacteria formation.
Warning: disposables must be compatible with the medicine delivered.
Warning: patient or operator injury may result if package is opened or
damaged, or if damaged components are used. Visually inspect contents and
package before each use.
Warning: air embolism can cause death or serious injury to the patient. Do
not connect a patient to the BodyGuard 595 Color Vision™ Infusion Pump
until all trapped air has been cleared from the fluid path. Carefully read the
instructions for loading. Puncturing set components may cause air embolism.
Warning: drugs must not be administered to the epidural space unless the
drugs are indicated for this purpose and are administered in accordance with
the indications included in the manufacture’s package insert accompanying
the drugs. Epidural administration of drugs other than those indicated for
epidural use could result in serious injury to the patient.
Caution: component damage may occur if not installed properly. Assure all
connections are secure; do not over tighten. This will help minimize leaks,
disconnection and component damage.