BodyGuard 595™Color Vision Operator Manual
To prime the set select the prime function from the
menu.
Select Change Bag and press .
Change the bag and press again.
Confirm protocol data and start infusion.
NOTE: if using an infusion set with drip chamber, change to a new infusion bag,
but make sure that the drip chamber is half filled and there is no air in the
administration set.
NOTE: if using an ambulatory administration set without a drip chamber, priming
the line might be needed. Disconnect the set from patient before priming.
Eliminating Air During Operation
When air is detected in the line, the pump stops the current infusion, sounds an audible
alarm, the screen shows Air or Up Occlusion, and the LED indicator turns from green to
red.
To eliminate air in the line during pump operation, follow these steps:
10:00 - 70% █
■
Prime
Select Protocol
Change Bag
View Protocol Settings
View Shift Totals
View Patient History
View Event Log
Clinician Bolus
Select , and press
10:00 - 70% █
■
Priming
██████████████
0 5mL
(Press key to Stop)
10:00 - 70% █
■
Change Bag
Press when done
10:00 - 70% █
■
Prime
Select Protocol
Change Bag
View Protocol Settings
View Shift Totals
View Patient History
View Event Log
Clinician Bolus
Select , and press