Depending upon the set flow rate and pressure
threshold, the Down Occlusion alarm will be
activated.
Press to mute the alarm.
Door Open – Alarm Test
Insert a MicroSet
TM
in the pump and close the door. Press the door until a click is
heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
Press until the Self-Test screen appears. If air sensor is off, press to
confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
Start infusion as instructed in the "Select protocol" section.
Door Open alarm will be activated at once.
Close the door and press .
Battery Test
Insert a MicroSet
TM
in the pump and close the door. Press the door until a click
is heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
Take the pump out of the charger.
REGIONAL 10:00 - 70% █
■
Down
Occlusion
Check Set / Access
Press key to mute
PCA 10:00 - 70% █
■
Door Open
Close door