EasyManuals Logo

CME BodyGuard 595 Color Vision User Manual

CME BodyGuard 595 Color Vision
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
4 Operation
BodyGuard 595Color Vision Operator Manual
24
7.
Depending upon the set flow rate and pressure
threshold, the Down Occlusion alarm will be
activated.
Press to mute the alarm.
Door Open Alarm Test
1.
Insert a MicroSet
TM
in the pump and close the door. Press the door until a click is
heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
2.
Press until the Self-Test screen appears. If air sensor is off, press to
confirm. Please contact your technician in order to enable the air sensor if
disabled with no need.
3.
Start infusion as instructed in the "Select protocol" section.
4.
Open the pump door.
5.
Door Open alarm will be activated at once.
6.
Close the door and press .
Battery Test
1.
Insert a MicroSet
TM
in the pump and close the door. Press the door until a click
is heard.
NOTE: for instructions on how to load and prime an administration set refer to
sections “Loading the Administration Set” & “Priming the Line”.
2.
Take the pump out of the charger.
REGIONAL 10:00 - 70%
Down
Occlusion
Check Set / Access
Press key to mute
PCA 10:00 - 70%
Door Open
Close door

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 595 Color Vision and is the answer not in the manual?

CME BodyGuard 595 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 595 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals