BodyGuard 595™Color Vision Operator Manual
Default Settings
The pump comes programmed with a number of default parameters that can be viewed
using ‘View Set Up’ from the main menu and changed using the technician access code
in ‘Change Set Up’ mode.
Amount of fluid delivered during
set priming.
Rate at which priming occurs.
Maximum bolus volume for a
protocol
Maximum bolus volume allowed by
a clinician
Rate at which boluses are
delivered
40-1200ml
System will adapt rate
according with in-line
backpressure
Maximum continuous basal rate
PCA 0.1 - 100 ml/hr
Epidural 0.1 - 30 ml/hr
Regional 0.1 - 100 ml/hr
A continuous rate after infusion
program has ended which is
designed to prevent coagulation
and keep an IV line open
0ml/hr (Disable)
0.1-5ml/hr
Minimum lockout time between
boluses
Switches the LED indicator
On & Off
Change bolus volume during
infusion
Number of seconds the info screen
is displayed before returning to the
current infusion screen
Change between available
language (may vary between
software version)
English, Spanish, Italian,
German (Deutch), French,
Dutch
Serial Number Unique Pump
Volumetric calibration value
Pressure calibration Delta value
Pressure calibration CAP value
Automatically test all pumps