4 Operation
BodyGuard 595Color Vision Operator Manual
54
program accumulation screen and during program operation.
Setting a protocol based on basal rate when the bolus is set to zero
Pump will enable program operation only if the basal rate set is equal or less than
maximal MediGuard
TM
rate. In case the program basal rate is less than MediGuard
TM
maximal allowed rate, the pump will count the difference between the two different rates.
The difference between the actual rate and the maximal MediGuard
TM
rate will be stored
in a buffer called “credit buffer . Maximum volume saved in the credit buffer is volume
of 10 minutes delivery of the MediGuard
TM
maximal rate. The credit is used to enable
delivering of clinician bolus. If Clinician bolus volume is less than the volume in the
buffer, the bolus will be delivered at maximal bolus rate. If Clinician bolus volume is
higher than the volume in the buffer, the volume in the buffer will be delivered at
maximal bolus rate and the remaining bolus volume will be delivered at MediGuard
TM
maximal rate (if there is no volume in the credit buffer all bolus will be delivered at
MediGuard
TM
maximal rate).
Setting a protocol based on a combination of boluses and basal rate (or only
boluses)
Pump will enable program operation only if the basal rate set is equal or less than
maximal MediGuard
TM
rate. In case the program basal rate is less than MediGuard
TM
maximal allowed rate, the pump will count the difference between the two different rates.
The difference between the actual rate and the maximal MediGuard
TM
rate will be stored
in the credit buffer. As program begins, the credit buffer will automatically include
volume for one bolus. Maximum volume saved in the credit buffer is the bolus volume.
The credit buffer enables delivery of patient and clinician boluses. When a bolus is
required (either by patient or by clinician): if the bolus volume is less than the volume in
the buffer, the bolus will be delivered at maximal bolus rate; if the bolus volume is higher
than the volume in the buffer, the volume in the buffer will be delivered at maximal bolus
rate and the remaining bolus volume will be delivered at MediGuard
TM
maximal rate.
NOTE: setting MediGuard
TM
option ON shall disable the Clinician bolus option
from the menu. Any attempt to give a clinician bolus from the Menu will display
the following message: MediGuard
TM
set ON Clinician boluses can be Given
from protocol only
NOTE: the MediGuard
TM
option is replacing the hour limit, therefore setting
MediGuard
TM
ON, the hours limit will be disabled and not shown in the
parameters settings.
Using History Functions
The BodyGuard 595 Color Vision keeps a rolling 24 hour history of the current
patient’s infusion delivery. In this section we will address how this can be used. There is
also a 2000 event history log keeping track of the last 2000 user actions in a date and
time stamped record. During infusion press the button repeatedly to scroll through
the following information screens:
Shift totals

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 595 Color Vision and is the answer not in the manual?

CME BodyGuard 595 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 595 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals