BodyGuard 595™Color Vision Operator Manual
• Current infusion totals
• Bolus summary
• Protocol information
• Battery level
• Time & date
• Pump information
Press to stop the infusion and then press repeatedly to scroll through the
following information screens:
• Volume and boluses given in the last 1 hour
• Bolus attempted and given and volume infused hour-by-hour for the last 24 hours
(use the arrow keys to scroll through the hours)
• Chart of bolus delivery during the last 24 hours, hour-by-hour
• Chart of volume (incl. boluses) delivered during the last 24 hours, hour-by-hour
Press to resume infusion when you return to the ‘STOP’ screen.
Events Log
To access the events history enter 'View Event Log' from the menu
The events history contains the last 2000 user events (i.e. changes to programs, pump
started, stopped, etc) and the status of the pump (i.e. battery charge, pressure levels,
etc). It is not patient specific (i.e. the 2000 events are likely to span many patients
recently treated with that particular pump).
To access the events history:
From the main menu scroll to ‘View Event Log’ and press
The screen will now show the most current user
event, date, time, etc.
NOTE: each event is assigned a new number and the pump stores the last 2000 in
memory. After some time the first event to appear when you enter the events
history may be number 2754. This means there have been 2754 events in this
pumps life and events 754-2754 are stored in this history. The pump deletes the
oldest event in the log each time a new event occurs.
Use the Up & Down arrow keys to scroll through events to find a specific event.
When you find one of interest:
Event Number: 123456
01-Jan-2000 10:00
Protocol Starts
Press for Details