BodyGuard 595™Color Vision Operator Manual
If auto bolus:
Press to commence infusion.
Press to delay auto-bolus.
Press to deliver a loading dose and start the
infusion.
Titrating Basal Rate During Infusion
To change Rate during the infusion, enter the new rate using the keypad and confirm by
pressing the key. Confirm again the new rate by entering the Relevant Code. The
Rate can be increased up to the “Max Titration Rate” programmed into the specific
infusion. The pump will beep to confirm acceptance of the change and the new rate will
be displayed.
NOTE: if the desired rate is above the protocol maximum range the pump will not
accept the entry and will display a message notifying the maximum rate allowed.
NOTE: the protocols allow individual setting of the following parameters:
Single air bubble
Pressure threshold
Bolus volume
Clinician Bolus
Patient Activated Bolus
The BodyGuard 595 Color Vision™ infusion pump can be programmed to allow the
patient to request boluses of a specified volume within predetermined time intervals.
The patient can request a bolus by one of the following two methods:
Using a Bolus Cable for various modes
SAFETY CHECK: Prior to connecting the bolus cable to the pump test the bolus cable.
Press and release the handset button – you should feel and hear an audible click each
time. Contact biomedical engineering or your local distributor in order to replace the
bolus cable.
Press the button located at the end of the bolus cable. The bolus cable plugs into the
side of the pump where a label reads “Bolus”.
The bolus button must be pressed for a minimum time in order to activate a bolus.
This feature prevents unintentional bolus.
The bolus button must also be released within a certain timeframe. This feature
prevents unintentional bolus in case a button is accidentally sat on or otherwise
permanently pressed.
EPIDURAL 10:00 - 70% █
■
Start Auto-Bolus?
Press key to start
Press to Delay Auto-Bolus
for Loading Dose & Start