BodyGuard 595™Color Vision Operator Manual
To Unlock Keypad
Press and hold until the entire graph is clear and a
beep is heard. The beep indicates that the locking
program is turned OFF.
NOTE: to prevent unauthorized individuals accessing the main menu a technique
is required to return to the menu code screen when the pump is in STOP mode.
To return to the menu during infusion, press key to stop delivery and enter
the STOP state.
Press and hold key until the level one access code prompt appears.
NOTE: when keypad lock is on, entering the STOP state & holding down
key will not return you to the main menu. Press to restart the
infusion, press and hold to remove keypad lock, then press and
hold key until you are prompted for the level one access code to return to
the main menu.
Dedicated Administration Sets
The BodyGuard 595 Color Vision™ Pain Manager should be operated only with
MicroSet
TM
dedicated Infusion sets, deferent color for deferent program: PCA (Blue),
Epidural (Yellow) and Regional (Green).
Warning: do not operate the BodyGuard 595 Color Vision™ Pain Manager
with any administration set, other than the MicroSet
TM
. The MicroSet
TM
infusion sets are equipped with a free flow protection valve, to prevent a
gravity free flow hazard. Using a different set may essentially change the
delivering rate and expose the patient to a free flow rate, which can cause
serious injury or death.
Warning: using the pump with air in line detector off, may cause an embolism
resulting death or paralysis. Switching the air detector off is not recommended
and should be used only when absolutely necessary. Please contact your
technician in order to enable the air sensor if disabled with no need.
EPIDURAL 10:00 - 70% █
■
Keypad
Lock Mode
███████████████
Off On