4 Operation
BodyGuard 595Color Vision Operator Manual
32
Modify Protocol
When you receive your BodyGuard 595 Color Vision pump and go to "Modify Protocol
Library" on the main menu for the first time you will see the protocols are blank. On the
first few patients using the BodyGuard 595 Color Vision you can edit, add or delete
any of these infusion protocols to tailor the protocols to your service.
This section is probably the most complex of the manual but remember you should only
have to perform this function once for each protocol with each pump whilst you are
setting it up for use on the first few patients. It is possible to adjust the pump’s default
protocols to the one you require, by request from our factory.
Each protocol contains the following data:
Infusion Mode (PCA, Epidural, Regional)
Drug/protocol name
Minimum & maximum patient weight
Volume to be infused
Initial rate in ml or mg or mcg per hour (Initial Rate and Protocol Maximum rate
are superseded by the values set in the 'Change Set up Menu), (If MediGuard
TM
is on Rate can only be entered based on body weight units)
Protocol Maximum Rate
Bolus dose and maximum bolus dose (if required)
Lockout time between boluses and maximum lockout time (if required)
Maximum clinician (CA) bolus
Bolus rate
Hour limit applicable to this regime (if required)
Keep Vein Open rate (if required)
Occlusion pressure
Air-in-line limit
If MediGuard
TM
is on maximum MediGuard
TM
rate
NOTE: you can use protocols to set up several commonly used treatment
regimes in your service i.e. same bag, drug name, basal rate but different bolus
volumes or lockout times.
NOTE: after setting up your most common regimes as a protocol set up the next
or last protocol letter as a Tailored’ or one off regime that you can modify for
individual patients who dont fall within the pre-set protocols. Once the infusion
has been administered for a particular patient you can go back in and change this
for the next or a future patient.
NOTE: each protocol parameters are unique to that protocol. For example if you
use both Ropivacaine and Bupivacaine in your service and one commonly comes
in a 100 ml bag whilst the other is in a 250 ml bag than you can set these bag
sizes as standard for each.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CME BodyGuard 595 Color Vision and is the answer not in the manual?

CME BodyGuard 595 Color Vision Specifications

General IconGeneral
BrandCME
ModelBodyGuard 595 Color Vision
CategoryMedical Equipment
LanguageEnglish

Related product manuals