BodyGuard 595™Color Vision Operator Manual
Press to silence the alarm. Press and hold down for 3 seconds to
return to the menu screen.
Enter the relevant Access Code.
Disconnect the line from the patient; use aseptic technique.
Warning: ensure the line is disconnected from
the patient before proceeding to the next step.
Failure to do so could result in the patient
receiving air through into the bloodstream.
Press within 10 seconds to commence priming, otherwise the display will
return to menu. You may stop priming at any time by pressing .
Repeat steps 3-6 until all air is eliminated from the line and fluid is seen at the
distal end of the set.
Press to return to the menu option screen.
Press to resume the current infusion.
Check the Infusion Summary if screen which shows: bag volume, volume left and
infused, is correct then press again to restart delivery
Down Occlusion Alarm
After Down Occlusion Alarm press .
REGIONAL 10:00 - 70% █
■
Warning !
Disconnect Patient
Press key to Prime
REGIONAL 10:00 - 70% █
■
Down
Occlusion
Check Set / Access
Press key to mute