4 Operation
BodyGuard 595Color Vision Operator Manual
42
40.
Set KVO rate and press . The to 0 if not
applicable.
41.
Set occlusion pressure alarm level specific to the
protocol/patient and press .
42.
Set air-in-line limit specific to the protocol/patient
and press .
43.
If the MediGuard
TM
feature is enabled, the pump
will ask for "Max MediGuard
TM
Rate" in ml, mg or
mcg per hour per kg bodyweight. This sets a
toxicity limit for the protocol based on the patient's
weight.
PCA 10:00 - 70%
A: MORPHINE
Keep Vein Open Rate
0.5 mL/hr
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Occlusion Pressure
280 mmHg
(250-520 mmHg)
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
MediGuard
On
Select , and press
PCA 10:00 - 70%
A: MORPHINE
Max MediGuard Rate
8 mg/kg/hr
Enter and Press Key
PCA 10:00 - 70%
A: MORPHINE
Air-in-line Limit
0.1 mL
Enter and Press Key