28
GenesisPlus Operator Manual
D1151, R
EV C, 10/16
System does not turn on. The touchscreen does not illuminate.
Probable Cause: The system is not plugged in.
Suggestion: Place the system main power circuit breaker in the OFF (down)
position, insert the system electrical power cord plug into the
appropriate outlet and place the system main power circuit
breaker in the ON (up) position. Turn the keyswitch to the
START position.
Probable Cause: The building power (main electrical service) is turned off.
Suggestion: Turn on the building power or reset the room circuit breaker.
Probable Cause: The system main power circuit breaker is in the OFF (down)
position.
Suggestion: Place the system main power circuit breaker in the ON (up)
position.
Probable Cause: The electrical outlet is defective.
Suggestion: Use another outlet or have the outlet professionally tested and,
if appropriate, repaired.
Probable Cause: The remote interlock plug is not connected to the system console.
Suggestion: Insert the remote interlock plug into the interlock receptacle, then turn
the keyswitch to the START position.
Treatment beam is not present, although aiming beam operates properly.
Probable Cause: There is an internal laser system failure.
Suggestion: Contact Cutera Service (refer to Appendix D).
Treatment and aiming beams are not present. Touchscreen displays and indicators are
normal.
Probable Cause: The system is in STANDBY mode.
Suggestion: Place the system in READY mode.
Probable Cause: There is an internal laser system failure.
Suggestion: Contact Cutera Service (refer to Appendix D).