Probable Cause: The system main power circuit breaker has tripped.
Suggestion: Place the system main power circuit breaker in the UP
position.
Probable Cause: Remote interlock is in use and the treatment room door
has been opened.
Suggestion: Close the treatment room door and restart the system.
“Water Low” advisory pop-up screen appears on the touchscreen.
Probable Cause: The system is low on water.
Suggestion: Turn off the system, and add water as instructed in the
Routine User Maintenance topic. Restart the system to
continue with treatments.
Low or no energy output (patient reports no sensation).
Probable Cause: Handpiece lens is dirty or damaged.
Suggestion: Confirm that laser energy is not being emitted on the test
area, such as on an arm. Check to see if the aiming beam is
present. Clean the lens as described in this manual. If the problem is
not resolved, contact Cutera Service (refer to Appendix D) for lens
replacement.