Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The GenesisPlus is intended for use in the electromagnetic environment specified below. The customer or the user of the
GenesisPlus should ensure that it is used in such environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment: Guidance
Electrostatic Dis-
charge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±2, 4, 6kV contact
±2, 4, 8kV air
Floor should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast tran-
sient/burst
IEC61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV line to ground
±kV line to line
Mains power quality should be that of a typ-
ical commercial or hospital environment.
Surge
IEC61000-4-5
±1kV differential mode
±2kV common mode
±0.5, 1kV
differential mode
±0.5, 1, 2kV
common mode
Mains power quality should be that of a typ-
ical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
<5% Ut (95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec.
Mains power quality should be that of a typ-
ical commercial or hospital environment. If
the user of the GenesisPlus requires contin-
ued operation during power mains interrup-
tions, it is recommended that the
GenesisPlus be powered from an uninter-
ruptible power supply or a battery.
Power frequency
(50/60Hz) mag-
netic field
IEC 61000-4-8
3 A/m N/A Power-frequency magnetic fields should be
at
levels characteristics of a typical location
in a typical commercial or hospital environ-
ment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
56
GenesisPlus Operator Manual
D1151, R
EV C, 10/16