Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The GenesisPlus is intended for use in the electromagnetic environment specified below. The customer or the user of the
GenesisPlus should ensure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment: Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150kHz to 80 MHz
3V/m
80MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment
should be used no closer to any part
of the GenesisPlus, including its cables, than
the recommended separation distance calcu-
lated from the equation applicable to the fre-
quency of the transmitter.
Recommended Separation Distance
d = 1.17 P
d = 1.17 P 80MHz to 800MHz
d = 2.33 P 800MHz to 2.5GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site sur-
vey
(a)
, should be less than the compliance level
in each frequency range
(b)
.
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE: 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles
radios,
amateur radio, AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered,. If the measured field strength in the location in which the GenesisPlus is used exceeds the applicable RF compli-
ance level above, the GenesisPlus should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the GenesisPlus.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
57
GenesisPlus Operator Manual
D1151, R
EV C, 10/16