Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the GenesisPlus
The GenesisPlus is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The user of the GenesisPlus can help prevent electromagnetic interference by maintaining a minimum distance between
portable
and mobile RF communications equipment (transmitters) and the GenesisPlus as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
d = 1.17 P
80 MHz to 800 MHz
d = 1.17 P
800 MHz to 2.5 GHz
d= 2.33 P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.37
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.