Operating Instructions RF-2000
38
(2) Press Enter, and the following is displayed.
EX(nm) EM(nm) Data
TIME FUNC
VALUE
Time: Elapsed time from time program start (minute)
Func: Command name
Value: Set value
(3) If a time program is made, the following is displayed by pressing Enter again.
The contents in the display are the same as those in (2).
(Example)
EX(nm) EM(nm) Data
<<1.00>> EX λ
λλ
λ 350
The above display indicates that the excitation wavelength becomes 350nm in one
minute from the time program start.
(4) When pressing Enter repeatedly, the time program is displayed in forward
steps.
(5) Pressing shift + back, the time program is displayed in backward steps.