Operating Instructions RF-2000
40
5.4.4 Scanning by program
Example setting to execute scanning in 5 minutes and store data in file No. 2 using a
program is shown below:
(1) Press CE to return to the initial screen.
EX(nm) EM(nm) Data
<<350>> 400
1000.00
(2) Press shift + edit.
EX(nm) EM(nm) Data
0 USED 32
LEFT
(3) Press Enter.
EX(nm) EM(nm) Data
<<TIME>> FUNC
VALUE
(4) Enter a time (minute). Press 5 and Enter.
EX(nm) EM(nm) Data
5.00 <<EX λ
λλ
λ>>
(5) Select a command. Press func repeatedly until SCAN is displayed.
Press Enter to set SCAN.
EX(nm) EM(nm) Data
5.00 SCAN
<<VALUE>>
(6) Enter a file No. where data is to be stored. Press 2 and Enter.
EX(nm) EM(nm) Data
<<5.00>> SCAN 2
(7) Enter the second and later steps if necessary by repeating (4) to (6). To finish
setting the program, press CE twice.
Please note:
When setting multiple steps, it is not necessary to set the
commands in order of time. They are automatically
rearranged.
Set the STOP command at the end of a program unless the
program is to be executed repeatedly.
In selecting a function, the previous function is displayed by
pressing shift + back.