Operating Instructions RF-2000
48
During scanning the excitation wavelength, this unit measures the spectrum without
the division correction, because the corrected spectrum may not indicate the
optimum excitation wavelength for analysis with this unit.
XE TIME
(Display of operating time of xenon lamp)
EX(nm) EM(nm) Data
XE TIME 100
This function displays accumulated operating time of xenon lamp (Unit: hour).
Press 0 and Enter to reset the accumulated time to 0.
Please note:
Do not reset the time except for when replacing the lamp.
Z WAVE
(Output holding function at change of wavelength, sensitivity or gain)
EX(nm) EM(nm) Data
Z WAVE __
This function sets operation/non-operation of output holding function on changing
wavelength, sensitivity or gain.
Set value Correction
0 Operation
1 Non-operation
Please note:
If the wavelength is changed, light energy in the background
changes and the output value varies. By setting this operation
(set value: 0), the output value does not vary when the
wavelength is changed. When changing the wavelength during
the analysis, set this operation.
This function can not operate correctly when response value is
set at "4-(3.0sec)".
RS232C
(Setting of RS-232C and starting communication)
EX(nm) EM(nm) Data
RS232C
Press the Enter key to display the commands for starting the communication and
setting parameters of RS-232C (see "7.2 Control from External equipment via RS-
232C").
CLOSE KEY
(Rejects any key input.)
EX(nm) EM(nm) Data
CLOSE KEY
This function refuses any key input.
After pressing Enter, any key input is rejected.