Service Manual
Ophthalmic Surgical System
Chapter 2 - Maintenance instructions
22 - Version date: Nov. 01, 2013 - N° doc.: 30204400-B
10. Remove the CAN
cable.
11. Unscrew the 4 self-locking nuts
12. Remove the drive unit carefully.
The drive unit is mounted on 4 vibration damp-
ers in which set screws are mounted with lock-
ing agent (Threebond normal 1374).
If a set screw is released from the damper, refit
it with locking agent.
The screws are mounted with
locking agent (Threebond normal
1374).
Assemble the screws with
Threebond.