Technical Data
108
Environmental conditions
Settings
Technical Data
Environmental conditions
In operation
Temperature 10 to 40
o
C
Atmospheric pressure 700 to 1060 hPa
Rel. humidity 0 to 90 %, without condensation
In storage
Temperature –20 to 60
o
C
Atmospheric pressure 500 to 1060 hPa
Rel. humidity 0 to 100 %, without condensation
Settings
Ventilation modes IPPV/IPPVAssist,
SIMV, SIMV/ASB
MMV, MMV/ASB
BIPAP/ASB
BIPAP
Assist
CPAP/ASB
Ventilation frequency f 0 to 100/min
Inspiration time T
insp 0.1 to 10 s
Inspiration time V
T
Paediatrics 0.02 to 0.3 L, BTPS
*
* BTPS = Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient lung: Body temperature 37
o
C, steam-saturated gas, ambient pressure.
Accuracy ±10 % of set value, or ±10 mL, whichever is greater.
Adults 0.1 to 2.0 L, BTPS*
Accuracy ±10 % of set value, or ±25 mL, whichever is greater.
Inspiratory Flow
Paediatrics 6 to 30 L/min
Adults 6 to 120 L/min
Inspiratory pressure P
insp 0 to 80 mbar
Inspiratory pressure limit Pmax 0 to 100 mbar
O
2 concentration 21 to 100 Vol.%
Accuracy ±5 % of set value, or ±2 Vol.%, whichever is greater.
Positive end-expiratory pressure PEEP or interm. PEEP 0 to 35 mbar
Trigger sensitivity 0.3 to 15 L/min
Pressure assist P
ASB 0 to 80 mbar
Rise time for pressure support (Ramp) 0 to 2 s
I:E 1:9.5 to 4:1 (can only be configured in default settings)