77
Configuration
Ventilation Defaults
Ventilation Defaults
— For defining the patient-specific start-up parameters »f«
and »VT«
— Ventilation parameters active on starting up the device
— Alarm limits active on starting up the device
— Ventilation mode
The »Ventilation« menu for the default settings active on
starting ventilation can only be accessed after entering the
code number 3032.
This code-protection is designed to prevent accidental
changes to the configuration.
Patient-specific defaults
for Adult or Paediatric mode
— To set the values of the »f« and »VT« parameters active on
starting up the device.
● Press the »Calib./Config.« menu key.
● Press the »Ventilation jj« menu key.
● Enter code number »3032«.
● The »Patient« menu is displayed.
Display (example):
● Select the »VT« screen field = turn the rotary knob.
Confirm = press the rotary knob.
● Set the desired value = turn the rotary knob.
Confirm = press the rotary knob.
Patient-specific defaults VT, f
The hospital-specific defaults for your hospital can be entered
in the table.
Patient mode Factory setting Hospital-specific setting
Tidal volume VT
mL
Ventilation frequency f
1/min
Tidal volume VT
mL
Ventilation frequency f
1/min
Paediatric 50 29
Adult 500 12
524525