Configuration
Ventilation Defaults
Ventilation Defaults
For defining the patient-specific start-up parameters »f«
and »VT«
Ventilation parameters active on starting up the device
Alarm limits active on starting up the device
Ventilation mode
The »Ventilation« menu for the default settings active on
starting ventilation can only be accessed after entering the
code number 3032.
This code-protection is designed to prevent accidental
changes to the configuration.
Patient-specific defaults
for Adult or Paediatric mode
To set the values of the »f« and »VT« parameters active on
starting up the device.
Press the »Calib./Config.« menu key.
Press the »Ventilation jj« menu key.
Enter code number »3032«.
The »Patient« menu is displayed.
Display (example):
Select the »VT« screen field = turn the rotary knob.
Confirm = press the rotary knob.
Set the desired value = turn the rotary knob.
Confirm = press the rotary knob.
Patient-specific defaults VT, f
The hospital-specific defaults for your hospital can be entered
in the table.
Patient mode Factory setting Hospital-specific setting
Tidal volume VT
mL
Ventilation frequency f
1/min
Tidal volume VT
mL
Ventilation frequency f
1/min
Paediatric 50 29
Adult 500 12