Configuration
72
System Settings
Configuration
System Settings
Adjusting the volume of the audible alarm
Press the »Calib./Config.« menu key.
Press the »Device jj« menu key. The »System« menu
appears.
Display (example):
Select the »Loudness« screen field = turn the rotary knob.
Activate = press the rotary knob.
Set the desired volume = turn the rotary knob.
Confirm = press the rotary knob.
After the setting has been confirmed, the alarm tone is
sounded once to enable you to judge the volume.
Setting the contrast
Not possible with all types of monitor
Press the »Calib./Config.« menu key.
Press the »Device jj« menu key. The »System« menu is
displayed.
Select the »Contrast« field on the screen = turn the rotary
knob.
Activate = press the rotary knob.
Setting the contrast = turn the rotary knob,
confirm = press the rotary knob.
The set contrast will now be activated.
Set the volume of the acoustic alarm to a sufficiently high
level so that the alarm can be heard!