43
Maintenance
Calibration with test-gas ampoules
Use the calibration flask (Part No. 68 03 407) – Observe the information in the
sensor data sheet.
Follow the instructions printed on the calibration flask and enclosed with the
test-gas ampoules.
For units without a pump:
1 Fit the adapter and calibration bottle to the Polytron 7000.
Break the test-gas ampoule inside the calibration flask.
Wait until the measured value has settled (see the operating instructions for the
sensor for the necessary waiting period). Then carry out calibration.
For units with a pump:
The pump adapter is already installed on the Polytron 7000.
Insert the adapter (Part No. 68 04 620) in the calibration flask – connect the hos-
es.
Break the test-gas ampoule inside the calibration flask and wait until the meas-
ured value has settled (see the operating instructions for the sensor for
the necessary waiting period). Then carry out calibration.
05023758_1.eps
OK
68 07 407
04923758_1.eps
OK
+
05123758_1.eps
OK
68 07 407
68 04 620
+