EasyManua.ls Logo

Dräger P3U - Calibrating the Unit

Dräger P3U
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
42
Calibrating the unit
Ensure that the sensor is warmed up before it is calibrated. See the sensor data
sheet for the warming-up time.
For some sensors (such as oxygen sensors) the function » zero-point calibration «
is simply a test of the sensor function. The zero point is not actually calibrated
since this is not necessary for these sensors.
For critical applications, the calibration intervals should be defined in accord-
ance with the recommendations in EN 50073
1)
, EN45544-4
2)
and national reg-
ulations.
Note the calibration sequence!
First check the zero point and calibrate it if necessary, immediately after this,
check the sensitivity and adjust it as necessary.
Never calibrate the sensitivity before calibrating the zero point.
Calibration cannot be carried out if the date and time are not set.
Setting the date and the time, page 65.
Calibration menu, page 56 to page 57
Zero gas and test gas: see the information in the sensor data sheet and on the
pages 105 to 108.
Calibrating with test gas
Use a test-gas cylinder with pressure-reduction valve (a stainless-steel pressure
reduction valve for aggressive gases). Observe the information in the sensor
data sheet.
1 Mount a calibration adapter Part No. 68 06 978 (with two hose connectors) on
the Polytron 7000.
Vent the test gas leaving the adapter into a fume cupboard or into the open air
(with a hose connected to the second connector on the calibration adapter).
1)
EN 50073 – Guidelines for selection, installation, use and maintenance of devices for the
detection and measurement of flammable gases and oxygen.
2)
EN 45544-4 – Electrical devices for the direct detection and direct concentration measure-
ment of toxic gases and vapours – Part 4: Guidelines for selection, installation, use and
maintenance.
Caution:
Test gas must not be inhaled. Risk to health! Care must be taken about the
risks which can arise when using test gas; hazard instructions and safety advice
must be observed.
For details, see appropriate DIN Safety Data Sheets.
04823758_1.eps
1
Polytron
OK

Table of Contents

Related product manuals