The menu » Information «
54
Submenu » Sensor «
The submenu » Sensor « contains the functions for interrogating the sensor status.
Sensor vitality
This function is active only if the Polytron 7000 transmitter is equipped with the sen-
sor diagnostic dongle, see page 29.
This function displays the remaining sensitivity of the sensor.
Select the menu items » Information «, » Sensor « and » Vitality « in this order,
pressing the » « key after each selection.
The current Sensor Vitality is displayed.
Dräger Safety recommends that the sensor is exchange when the Sensor Vitality
value is less than 25.
Last calibration date
This function displays the date of the last calibration.
Select the menu items » Information «, » Sensor « and » Last cal. date « in this or-
der, pressing the » « key after each selection.
The date of the last calibration is displayed.
Next calibration date
This function displays the date on which calibration is next due.
Select the menu items » Information «, » Sensor « and » Next cal. date « in this or-
der, pressing the » « key after each selection.
The date on which the sensor is next due for calibration is displayed.
Sensor temperature
This function displays the current sensor temperature and the maximum sensor
temperature which has been recorded.
Select the menu items » Information «, » Sensor « and » Show sensortemp. « in
this order, pressing the » « key after each selection.
The current sensor temperature and the maximum sensor temperature which has
been recorded are displayed.