The menu » Information «
54
Submenu » Sensor «
The submenu » Sensor « contains the functions for interrogating the sensor status.
Sensor vitality
This function is active only if the Polytron 7000 transmitter is equipped with the sen-
sor diagnostic dongle, see page 29.
— This function displays the remaining sensitivity of the sensor.
● Select the menu items » Information «, » Sensor « and » Vitality « in this order,
pressing the » « key after each selection.
— The current Sensor Vitality is displayed.
— Dräger Safety recommends that the sensor is exchange when the Sensor Vitality
value is less than 25.
Last calibration date
— This function displays the date of the last calibration.
● Select the menu items » Information «, » Sensor « and » Last cal. date « in this or-
der, pressing the » « key after each selection.
— The date of the last calibration is displayed.
Next calibration date
— This function displays the date on which calibration is next due.
● Select the menu items » Information «, » Sensor « and » Next cal. date « in this or-
der, pressing the » « key after each selection.
— The date on which the sensor is next due for calibration is displayed.
Sensor temperature
— This function displays the current sensor temperature and the maximum sensor
temperature which has been recorded.
● Select the menu items » Information «, » Sensor « and » Show sensortemp. « in
this order, pressing the » « key after each selection.
— The current sensor temperature and the maximum sensor temperature which has
been recorded are displayed.