The menu » Settings «
74
● Use the » « and » « keys to select » Enable « and press the » « key to ac-
tivate.
— The current for the warning signal is transmitted on the analogue interface.
● Use the » « and » « keys to select » Disable « and press the » « key to
activate.
— The current for the warning signal is switched off.
– Test maintenance
— This function is used to set the analogue interface to the maintenance
signal » «.
● Select the menu items » Settings «, » Communication «, » Analogue interface «
and » Test maintenance « in this order, pressing the » « key after each selec-
tion.
● Use the » « and » « keys to select » Enable « and press the » « key to ac-
tivate.
— The current for the maintenance signal is transmitted on the analogue interface.
● Use the » « and » « keys to select » Disable « and press the » « key to
activate.
— The current for the maintenance signal is switched off.
LON Interface
This function group contains the functions for the LON interface. For details of oper-
ation with LON, see also “Polytron 7000 Operation via LON” on page 81.
– Neuron ID
Every Polytron 7000 has a Neuron ID. The Neuron ID uniquely defines the Neuron
chip on the LON network.
To display the Neuron ID:
● Select the menu items » Settings «, » Communication «, » LON Interface « and
» Neuron ID «, pressing the » « key after each selection.
— The Neuron ID is displayed.
– Service PIN
The Polytron 7000 can be commissioned by sending its Neuron ID to the LON net-
work with the aid of the function "Service PIN".
● Select the menu items » Settings «, » Communication «, » LON Interface « and
» Service PIN « pressing the » « key after each selection.
● Use the » « and » « to select » Confirm « and press the » « key to activate
this selection.
— The Neuron ID of the Polytron 7000 is transmitted.
Note:
These functions may trigger alarms in the control unit! If necessary, the
alarms should be disabled in the control unit before using the functions.