EasyManua.ls Logo

Dräger P3U - Page 87

Dräger P3U
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
87
Technical Data
IECEx
Type P3U: EEx ia IIC T4 (–40
o
C Ta +65
o
C)
EEx ia IIC T6 (–40
o
C Ta +40
o
C)
IECEx BVS 04 0003 X
Power Supply: U
i
= 30 V, I
i
= 0,3 A, P
i
= 700 mW, C
i
= 5 nF, L
i
= 50 μH
Year of construction (via serial number)
1)
Dräger Safety, 23560 Lübeck, Germany
UL (Underwriters Laboratories Inc.)
Type P3U: Only as to Intrinsic Safety for use in Hazardous Locations
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Use in accordance with Dräger Control Drawing SE20105.
T4: –40 Ta +65
o
C, T6: –40 Ta +40
o
C.
Not tested in oxygen enriched atmospheres (>21 % O
2
).
Power Supply: V
max
= 30 V, I
max
= 0.3 A, P
i
= 700 mW,
C
i
= 5 nF, L
i
= 50 μH
CSA (Canadian Standards Association)
Type P3U: Intrinsic safe
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Ex ia T4 (–40
o
C Ta +65
o
C),
Ex ia T6 (–40
o
C Ta +40
o
C)
Use in accordance with Dräger Control Drawing SE20106.
Power supply: V
max
= 30 V, I
max
= 0.3 A, P
max
= 700 mW,
C
i
= 5 nF, L
i
= 50 μH
NRTL/C
®
LR97594

Table of Contents