DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual EMC Information
- 212 -
observed, additional measures may be necessary, such as reorienting or relocating the DUS 60
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Recommended separation distances between portable and mobile
RF communication equipment and the EQUIPMENT or SYSTEM-
For EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications
equipment and the DUS 60
The DUS 60 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the DUS 60 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the DUS 60 as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated maximum
output power of
transmitter
(W)
150 kHz to 80 MHz
P
V
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5.3
80 MHz to 800 MHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5.3
800 MHz to 2.5 GHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
7
0.01 0.35 0.35 0.7
0.1 1.1 1.11 2.21
1 3.5 3.5 7
10 11 11.1 22.1
100 35 35 70
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Order List
- 213 -
Appendix VI: Order List
The following accessories are recommended to be used on the DUS 60.
WARNING
Probes and other accessories used on the DUS 60 must be provided or recommended
by EDAN. Otherwise, the device may be damaged.
Part Name Part Number
Probe C363UA 12.01.116212
Probe L743UA 12.01.116214
Probe L742UA 12.01.116220
Probe L763UA 12.01.116222
Probe C343UA 12.01.116213
Probe C362UA 12.01.116217
Probe C321UA 12.01.116221
Probe C613UA 12.01.116215
Probe E613UA 12.01.116216
Probe E743UA 12.01.116223
Rechargeable Lithium-Ion battery 01.21.064135
Coupling gel 11.57.78001
Freeze footswitch 11.10.102414
Mobile trolley 03.28.328004
Hand carried bag 01.56.465013
Video printer (SONY UP-897MD) 11.18.520146
Video printer (MITSUBISHI P93W) 11.18.52147