DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Ultrasound Intensity and Safety
- 160 -
Acoustic Output Reporting Table
Transducer Model: C613UA
Operating Model: PW Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
≤1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.4034 0.4474 1.5775 0.5415
P
r.3
MPa 0.9316
W
0
mW 17.62 17.62 17.62
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 1.4 1.4
deq(Z
sp
) (cm) 0.0144
f
c
(MHz) 5.3316 5.3316 5.3316 5.3316
X(cm) 0.896 0.896 0.896
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.58 0.58 0.58
PD (usec) 1.7398
PRF (Hz) 6361
P
r
@PII
max
(MPa) 1.3869
d
eq
@PII
max
(cm) 0.0144 0.0144
FL
x
(cm) 0.469 0.469 Focal
Length
Fl
y
(cm) 3.8273 3.8273
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.053
Control
Conditions
Control1
AP=15; Frequency=5.5MHz;
Depth=29mm; Focus=10mm;
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Ultrasound Intensity and Safety
- 161 -
A2.3.3 : Test of Probe C343UA:
Test Item B B+M PW
p-, MPa 2.381 2.381 1.244
I
spta,
mW/cm
2
21.6333 54.3890 1677
System settings Control1 Control1 Control1
Z
p,
mm 48.5 48.5 52
W
pb6, (
║), mm
(┴), mm
0.2901
0.3493
0.2901
0.3493
0.2878
0.2957
prr, kHz -- -- 5112
srr, Hz 37 27 --
Output beam Dimensions, cm
2
3.024 3.024 3.024
f
awf,
MHz 2.9572 2.9572 2.7031
APF, % -- -- --
AIF, % -- -- --
Maximum power, mW 39.97 34.47 82.8
I
ob,
mW/cm
2
13.2176 11.3988 27.381
Power-up mode B mode B mode B mode
Initialization mode
B mode B mode B mode
Acoustic output freeze
Yes Yes Yes
Z
tt
(mm) -- -- --
Z
ts
(mm) contact
contact contact
Inclusive modes -- -- --
Control1: AP=15; Frequency=2.0MHz; Depth=98mm; Focus=60mm;