DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Ultrasound Intensity and Safety
- 158 -
Acoustic Output Reporting Table
Transducer Model: C613UA
Operating Model: B Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8883 0.0725 0.1634
P
r.3
MPa 2.155
W
0
mW 5.316 5.316
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 1.3
deq(Z
sp
) (cm)
f
c
(MHz) 5.8865 5.8865 5.8865
X(cm) 0.896 0.896
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.58 0.58
PD (usec) 0.274
PRF (Hz) 2342
P
r
@PII
max
(MPa) 2.4022
d
eq
@PII
max
(cm) 0.0479
FL
x
(cm) 0.5431 0.5431Focal
Length
Fl
y
(cm) 4.4323 4.4323
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.2133
Control
Conditions
Control1
AP=15; Frequency=4.5MHz;
Depth=29mm; Focus=10mm;
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Ultrasound Intensity and Safety
- 159 -
Acoustic Output Reporting Table
Transducer Model: C613UA
Operating Model: B+M Mode
TIS TIB
Non-Scan
Index Label MI
Scan
A
aprt
1 A
aprt
>1
Non-scan
TIC
Global Maximum Index Value 0.8883 0.0655 0.0082 0.0302 0.1578
P
r.3
MPa 2.155
W
0
mW 4.799 0.336 0.336 5.135
Min
of[W
.3
(z1),(mW)
I
ta.3
(z1)]
Z
1
(cm)
Z
bp
(cm)
Z
sp
(cm) 1.3 1.3
deq(Z
sp
) (cm) 0.7584
f
c
(MHz) 5.8865 5.8865 5.8865 5.8865 5.8865
X(cm) 0.896 0.896 0.896 0.896
Associated
Acoustic
Parameters
Dim of
Aaprt
Y (cm)
0.58 0.58 0.58 0.58
PD (usec) 0.274
PRF (Hz) 2114
P
r
@PII
max
(MPa) 2.4022
d
eq
@PII
max
(cm) 0.0531 0.7584
FL
x
(cm) 0.5431 0.5431 0.896 Focal
Length
Fl
y
(cm) 4.4323 4.4323 0.58
Other
Information
I
pa.3
@MI
max
(W/cm
2
) 0.2133
Control
Conditions
Control1
AP=15; Frequency=4.5MHz;
Depth=29mm; Focus=10mm;