DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual System Control
- 46 -
Date Set freely Set the system date.
Time Set freely Set the system time, format: H/M/S.
SnapShot
Device
USB-Disk/Fixed Set the storage device of snap shot.
System
Sleep
√/ Null
Select whether the device enters
sleep mode when no operation is
performed for certain minutes.
Waiting time Set freely
Set the system waiting time to enter
sleep mode (5-60 min).
Refresh
Speed
1~10
Set the grade of glint speed of system
dormancy.
F1
None, save frame, save cine, save AVI, file
manager, sweep speed, PW invert, wall filter.
Define the F1 key, select one of the
pull-down options.
F2
None, save frame, save cine, save AVI, file
manager, sweep speed, PW invert, wall filter.
Define the F2 key, select one of the
pull-down options.
F3
None, save frame, save cine, save AVI, file
manager, sweep speed, PW invert, wall filter.
Define the F3 key, select one of the
pull-down options.
TI
TIS, TIB, TIC
Select the application tissue of
thermal index.
Table 5-4 General Presetting Information
Printer type Printer model
DJ 3600 HP DeskJet D2368
DJ Generic VIP HP DeskJet D2568, HP DeskJet D5568
HP DeskJet F2418, HP DeskJet F2488
HP Deskjet 2050, HP Deskjet 1050
HP Deskjet lnk Advantage 2010
LJ Mono HP LaserJet P2015, HP LaserJet P2035
Table 5-5 Presetting a Report Printer
You must restart the system to validate the change, including Language, Keyboard Sound, and
Report Printer. After you perform those presetting, and press Return, the system displays a
confirm dialog box to prompt you whether to restart the system.
5.7.4. Presetting Examination
Examination types include abdomen, obstetrics, small parts, gynecology, orthopedics, cardiology,
urology and vascular.
Take obstetric examination presetting for example, in the preset menu, move the cursor to
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual System Control
- 47 -
highlight Obstetric and press Set to display obstetric examination presetting dialog box.
Parameter 1 Tab
Figure 5-13 Obstetric Presetting -- Parameter 1 Tab
Item Setting Allows you to
Display B, M, B+B, 4B, B+M, PW Set display mode type.
Depth 19 mm ~ 245 mm (C363UA) Set examine depth.
A Power 0~15 Set acoustic power, 16 levels.
Focus Num 1/2/3/4 Set the number of focuses.
Scan Density High density/high frequency Set scanning density.
Scan Angle 0/1/2/3 Set scanning angle.
THI
Tissue harmonic imaging /
General
Set THI.
Line Persist 0~7 Set image line correlation.
H Reverse
√/ Null
Set the attribute of Horizontal reversal.
Probe
Display all the probe type
this device supports
Set the probe type to use.
Freq F1/F2/F3/F4/F5 Set the frequency of probe.
Post Process Gray map (L, A, B, C, D, S) Select a default gray map.
IP 0~7 Set the image parameter.
M/D speed 0/1/2/3 Set the M mode or D mode sweeping speed.
B Gain 0~130 Set the gain of 2D image, in 2 increments.
TSI General/muscle/fatty/fluid Set the type of TSI.
ATO 0/1/2/3 Set the degree of auto tissue optimization
V Reverse
√/ Null
Set the attribute of Vertical reversal.
Table 5-6 Obstetric Presetting Information – Parameter 1