DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Specifications
- 148 -
A1.6: General Technical Specifications
Display Modes B, B+B, 4B, B+M, M, and PW
Image Gray Scale 256 levels
Image Magnification
In area
Real time: ×1.0, ×1.44, ×1.96, ×2.56, ×4.0, ×5.76, ×9.0, ×16.0
Frozen (only available in B mode): ×1.0, ×1.78, ×4.0, ×16.0
Storage 504 MB
Cine Review 256 frames
Depth Adjustment Adjustable in real time in B, B+B, 4B, B+M, M, and PW modes
Frame Correlation
Coefficient
8 levels to adjust (0~7), (B, B+B, 4B, B+M modes, ineffective when
freezing)
Image Conversion
Up/Down flip, Left/Right flip,
90°rotate
Language Conversion
Chinese, English, etc. (The language options varies with language
software installed.)
Focus Position 16 levels to adjust
Focus Number Max. 4
Software Packages
Abdomen, obstetric, small parts, gynecology, orthopedics, cardiology,
urology, and vascular
B mode Measurement
Distance, circumference, area, volume, ratio, % stenosis, histogram and
angle
M mode Measurement Distance, time, slope, and heart rate
D mode Measurement Time, heart rate, velocity, acceleration, trace, and RI
Annotations
Patient name, age, sex, time, date, hospital name, doctor name, comment
(full-screen character editing)
Body Mark 130 types
USB Port USB 2.0
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Specifications
- 149 -
A1.7: Probe Specifications
This device can detect the probe automatically.
Model Application
C363UA/C362UA Abdomen, Gynecology, Fetal / Obstetrics, and Pediatrics
C343UA Abdomen, Gynecology, Fetal / Obstetrics, and Pediatrics
C321UA Abdomen, Gynecology, Fetal / Obstetrics, Pediatrics and Cardiology
C613UA Abdomen, Gynecology, Fetal / Obstetrics, Pediatrics and Cardiology
E613UA (Transvaginal / Transrectal): Gynecology, Fetal / Obstetrics, and Urology
E743UA (Transrectal): Rectum and the surrounding viscera, uterus, ovary and prostate
L743UA/L742UA
/L763UA
Small parts (galactophore, thyroid gland, prostate), Neonatal Cephalic, Peripheral
Vascular, Musculo-skeletal (both Conventional and Superficial)
NOTE:
Maximum transducer temperature rise during use: Less than 10 °C.
Expanded uncertainty of temperature test: U=0.4°C, k=2.
Specifications
Standard
Probe
Optional Probe
R60
convex
array
transducer
L43 High
frequency
linear array
transducer,
Endorectal
transducer
R20
Micro-
convex
array
transducer
R10
Endocavit
y
transducer
R40
convex
array
transducer
R10
Micro-
convex
array
transducer
L63
High
frequency
linear
array
transducer
R60
convex
array
transducer
L43
High
frequency
linear
array
transducer
Probe
C363UA
L743UA/
E743UA
C321UA E613UA C343UA C613UA L763UA C362UA L742UA
Central
frequency
3.5 MHz 7.5 MHz 3.5 MHz 6.5 MHz 3.5 MHz 6.5 MHz 7.5 MHz 3.5 MHz 7.5 MHz
Elements
number
96 96 80 96 96 96 96 128 128
Space of
elements
0.66mm 0.42mm 0.40mm 0.34mm 0.63mm 0.28mm 0.63mm 0.50mm 0.30mm
Max.
number of
active
elements
32 32 32 32 32 32 32 32 32