DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Operation
- 60 -
The system-defined examinations are abbreviated as follows:
Abd: Abdominal; OB: Obstetric; Sml: Small Parts; Gyn: Gynecology; Ortho: Orthopedics
Urol: Urology; Vas: Vascular
6.6.1. Generic Measurements in B Mode
The default measurement of B mode is distance measurement. B mode measurement menus are
shown as follows:
Figure 6-3 B Mode Generic Measurement and Calculation Menu
Distance
To measure distance:
1. Press Measure to activate measurement function.
2. Roll the trackball to highlight Distance, press Set to activate a
measurement cursor “+” on the screen.
3. Roll the trackball and press Set to anchor the start point.
4. Roll the trackball and press Set to anchor the end point.
5. Roll the trackball and press Set to begin a new distance measurement.
You can measure a maximum of four groups of data. The outcome
will be displayed in the measurement result window, as shown below.
6. Press Measure to finish and exit.
Dist1
Figure 6-4 Distance Measurement and the Results
DUS 60 Digital Ultrasonic Diagnostic Imaging System User Manual Operation
- 61 -
Circumference/Area
z Ellipse Method
To measure Circumference / Area:
1. Press Measure to activate measurement function.
2. Roll the trackball to highlight Cir/Area. Then select Ellipse, and
press Set to activate a measurement cursor “+” on the screen.
3. Roll the trackball and press Set to anchor the start point of fixed
axis of ellipse.
4. Roll the trackball and press Set to anchor the end point of fixed
axis of ellipse.
5. Roll the trackball, and press Set to define the size of the ellipse.
6. Roll the trackball and press Set to begin a new
circumference/area measurement. You can measure a maximum
of four groups of data. The outcome will be displayed in the
measurement result window, as shown below.
7. Press Measure to finish and exit.
Ellipse1
LAxis
SAis
Circ
Area
Figure 6-5 Ellipse Circumference/Area Method and the Results
z Trace Method
To measure Circumference / Area:
1. Press Measure to activate measurement function.
2. Roll the trackball to highlight Cir/Area. Then select Trace, and
press Set to activate a measurement cursor on the screen.
3. Roll the trackball and press Set to anchor the start point.
4. Roll the trackball to outline the region of interest. As you move the