© Edwards Limited 2015. All rights reserved. Page 29
Edwards and the Edwards logo are trademarks of Edwards Limited.
Maintenance
A352-03-880 Issue B
5.12.6 The pumping speed is poor
The connecting pipelines are too small in diameter.
The connecting pipelines are too long.
The inlet filter is blocked.
The pump is not suitable for the application.
The inlet intake screen is clogged.
The optional inlet filter is restricted.
The oil separating cartridge is clogged.
5.12.7 There is an external oil leak
The oil pump shaft seal is worn or damaged.
There is an oil leak from the gas ballast control.
There is an oil leak from the drain tap.
There is an oil leak from the oil filter O-ring.
5.12.8 The motor cuts out
Mains power supply voltage or frequency is incompatible with motor characteristics.
Check the circuit breaker setting.
Check the electrical connections.
The oil is too viscous or the room temperature is too low.
A pressure rise in the exhaust pipework.
The oil separating cartridge is clogged.
The pump exhaust line is restricted or blocked.
5.12.9 Pump emits fumes at exhaust or oil consumption is high
The separating cartridge is installed incorrectly.
The separating cartridge is broken or burst.
The automatic trap for re-injecting oil is blocked or its float is filled with oil.
The oil filler cap not properly tightened.
The pump is overheating.
The pump is filled with the wrong type of oil.
5.12.10 Oil back-streams to vacuum chamber
The integrated inlet valve is faulty.
The gas ballast is open.