SCU-1600 Control Unit for Turbomolecular Pump
March 10 121 Issue 1-a
7
MAINTENANCE
Referred
section
3.6.1
3.9.1
3.6.1
Countermeasures
Turn OFF, and connect the
STP motor connection cable
correctly. Perform tuning after
power ON operation.
When the error is displayed
after tuning, contact Service
office.
Contact Service office.
1. Connect the cable
correctly. If the cable is not
locked, lock it securely.
2. Contact Service office.
Contact Service office.
Contact Service office.
Contact Service office.
Contact Service office.
Contact Service office.
Contact Service office.
Probable causes
1. Incorrect connection of the
STP motor connection
cable.
2. Failure of the STP motor
connection cable.
3. Failure of the STP pump.
4. Failure of the SCU-1600.
Failure of the SCU-1600.
1. Incorrect connection of the
STP connection cable.
2. Failure of the STP
connection cable.
Failure of the SCU-1600.
Failure of the SCU-1600.
Failure of the SCU-1600.
Failure of the STP pump.
Failure of the SCU-1600.
Failure of the SCU-1600.
Pump operation
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the touch down
bearing.
Rotation is
inoperative.
Breaker OFF.
Decelerate and stop.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the touch down
bearin
.
Decelerate and stop.
Decelerate and stop.
Decelerate and stop.
Decelerate and stop.
Decelerate and stop.
Decelerate and stop.
Occurrence condition
Motor resistance
measurement is
impossible.
Rotational speed exceeds
rated speed.
Disconnection of the STP
connection cable.
Failure of record in
memor
.
Failure of serial
communication.
Abnormal power circuit.
Failure of record in
memor
.
Abnormal memory.
Failure of serial
communication.
LCD Messages
Motor Resistor Lost
Overspeed 1
Overspeed 2
Overspeed 3
P.Cable
Disconnected
PCB Record Failure
PCB->DSP Com
Fail
Power Supply
Failure
Pump Record
Failure
Ram Error
R-Unit Com Failure
Table 42 - Error messages after self test (continued)