Pos: 3.58 /0025 Üb erschriften/E LTRA 1.1 Übersc hriften/11 Dust or soot from th e resistance fur nace @ 9\m od_142562662 1251_9.doc x @ 75164 @ 2 @ 1
3.29 Dust or soot from the resistance furnace
Pos: 3.59 /0010 E LTRA/0015 s ervice_instructio ns/CS-2000_s ervice/05Faults_C S-2000/0 660 Modul Dust or s oot from the r esistance furn ace @ 9\mod_ 142554150414 4_9.docx @ 74 477 @ @ 1
Fig
49
: Front CS-2000
The sample has been inserted into the furnace too quickly:
See Operation Manual chapter Applications / Coal analysis”.
The sample weight is too high:
If a sample burns explosively, then there could be a shortage of oxygen. This can cause
some soot or oil vapors that pass through the hot zone of the furnace without burning. It
is necessary to reduce the weight of the sample, e.g. to 100mg or even to 50mg. If the
weight reduction doesn’t help, then operate differently by shifting the boat slowly towards
the beginning of the hot zone of the furnace. Leave the boat there and allow the sample
to start burning. Towards the end of the combustion, move the boat to the middle of the
hot zone, in order to obtain a complete combustion. See Operation Manual Applications
/ Oil analysis”.
The sample is metallic:
Perform analysis by using the induction furnace. See Operation Manual.

Table of Contents

Other manuals for ELTRA CS-2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA CS-2000 and is the answer not in the manual?

ELTRA CS-2000 Specifications

General IconGeneral
BrandELTRA
ModelCS-2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals