99
Reference Manual
00809-0100-4772, Rev FA
Troubleshooting
March 2016
Troubleshooting
Figure 7-12. Sensor Installation
A. Proper alignment
B. Improper alignment
C. Top view of sensor cavity
D. Side view of sensor cavity
E. Sensor cavity outer wall
F. Sensor cavity inner wall
G. Sensor
H. Sensor centerline is properly aligned with flowmeter centerline.
I. Sensor centerline is improperly aligned with flowmeter centerline: Damage to sensor will occur.
J. Sensor post
3. The sensor should remain as close to vertical as possible when applying force to
seat it. See Figure 7-13.
Figure 7-13. Sensor Installation—Applying Force
A. Apply force with hand until sensor is seated.
B. Sensor centerline must be aligned with flowmeter centerline
C. Sensor properly seated